1. Turn on PH meter
2. Disolve each buffer solution in 250ML of distilled water
3. Immerse the electrode into the PH 6.86 solution,(under the temperature of 25)
4. Press the “CAL”(calibration)button for 5 seconds and release
-Display will start flashing 6.86
-Wait until the display stops flashing.
Rinse the electrode with distilled water and dry it with filter paper.
1. Immerse the electrode in PH 4.00 solution,
-Press“CAL”button for 5 seconds, then press and release immediately second time
-Reading will start flashing 4.00
-Wait until display stops flashing
Rinse and dry the electrode with distilled water as before.
6. Place the meter back to the buffer solution to test calibration. If incorrect, please repeat the calibration
7. If the estimated pH of your test solution is known to be greater than 7.0, you should calibrate the meter with a solution of 6.86 and 9.18. Follow the calibration chart for this. The third step is the calibration method for 9.18 points.
package: One pH meter Bags Buffer Powder Instruction manual